<< Contents << Other Topics << Specs and Limits << GoLogicPro Specs and Limits
 
GoLogicPro Timer / Counter Limits
 
The TriggerForms provide 32-bit timers and counters. The counters can be from 3 to 232...
 
232 = 4,294,967,296
 
The maximum TriggerForm timer depends on the active sampling rate.
 
Max Timer at 500 MHz or faster
 
Timer resolution at 500 MHz or faster = 2 nanoseconds
 
= 4,294,967,296 x 2 ns
 
= 8,589,934,592 ns
 
= 8.59 seconds
 
Max Timer at 100 MHz
 
Timer resolution matches the sampling rate when sampling below 500 MHz.
 
The timer resolution at 100 MHz = 10 ns
 
= 4,294,967,296 x 10 ns
 
= 42,949,672,960 ns
 
= 43 seconds
 
State Analysis: "1 to 8 input clock"
 
72 channels: 2 ns timer resolution (500 MHz sample rate)
 
36 channels: 1 ns timer resolution (1 GHz sample rate)
 
18 channels: 0.5 ns timer resolution (2 GHz sample rate)
 
State Analysis: "clock on A16/etc..."
 
The maximum counter is 4.2 billion.
 
Timers are not available.
 